Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Non-volatile matter determination, also known as evaporation residue testing, is a fundamental purity assay that quantifies the residue left behind after a chemical reagent is evaporated and dried under strictly controlled conditions. This article explains the technical principle, applicable standards such as GB/T 9740, ISO 6353, and ASTM D1353, typical sample types, step-by-step gravimetric procedures, calculation methods, and industrial applications of non-volatile matter testing for chemical reagents.

Technical Overview

Non-volatile matter, frequently designated as evaporation residue or dry residue after evaporation, is the mass of material remaining after a measured quantity of a volatile chemical reagent is evaporated to dryness and the residue is dried to constant mass under specified conditions. Because every impurity that does not volatilize with the matrix — dissolved inorganic salts, suspended particulates, high-boiling organic by-products, stabilizers, and container leachables — is retained in the weighing vessel, the test functions as a broad-spectrum purity indicator for volatile reagents.

The determination is standardized internationally. GB/T 9740-2008 specifies the general method for the determination of dry residue after evaporation of chemical reagents in China and is harmonized in principle with ISO 6353-1. In the United States, ASTM D1353 covers nonvolatile matter in volatile solvents used in paints, varnishes, and lacquers, while the JIS K 8000 series in Japan provides equivalent general rules for reagent testing. Product monographs for individual reagents then assign numerical limits, typically expressed as milligrams per 100 milliliters or as a mass fraction in percent or milligrams per kilogram.

  • GB/T 9740-2008: General method for the determination of dry residue after evaporation of chemical reagents
  • ISO 6353-1: General method for the determination of dry residue after evaporation
  • ASTM D1353: Nonvolatile matter in volatile solvents for paints, varnishes, lacquers, and related products
  • JIS K 8000 series: General rules for test methods and reagents
  • GB/T 619: Sampling and acceptance rules for chemical reagents

Testing Samples

Organic Solvents

Methanol, ethanol, isopropanol, acetone, ethyl acetate, butyl acetate, toluene, xylene, hexane, and petroleum ether constitute the majority of routine samples. These solvents evaporate cleanly at water-bath temperature and are the simplest matrices for gravimetric residue determination.

Volatile Inorganic Acids and Alkalis

Hydrochloric acid, nitric acid, acetic acid, formic acid, and ammonia solution are regularly tested for evaporation residue. Hydrofluoric acid and fluoride-containing etchants require platinum vessels because they attack glass and quartz, and their corrosive fumes must be fully exhausted during evaporation.

High-Purity and Electronic-Grade Chemicals

Solvents, etchants, and cleaning agents used in semiconductor, display, and photovoltaic manufacturing are controlled at microgram-per-liter levels. Such samples demand clean-bench handling, rigorously controlled blanks, and complementary instrumental techniques such as ICP-MS and ion chromatography to identify the composition of the residue.

Specialty and Custom Samples

Custom formulations, diluted working reagents, recovered solvents, and rinsates are frequently submitted to accredited third-party laboratories for incoming inspection, lot-release verification, or failure investigation.

Testing Items

The core deliverable of the test is a quantitative statement of residue content, supported by methodological quality controls:

  • Non-volatile matter (evaporation residue) dried at 105 °C ± 2 °C
  • Result expression as mg/100 mL, mg/mL, percent by mass, or mg/kg (ppm)
  • Monograph-specific drying at alternative temperatures where the product standard requires
  • Vessel blank correction, duplicate determination, and constant-mass verification

Related gravimetric purity items frequently tested alongside non-volatile matter include ignition (sulfated) residue per GB/T 9741, loss on drying, water content by Karl Fischer titration, and acidity or alkalinity titration.

Testing Methods

Principle

A known volume or mass of sample is placed in a pre-weighed evaporating vessel, evaporated gently on a boiling water bath, dried in an oven at 105 °C ± 2 °C to constant mass, cooled in a desiccator, and re-weighed. The mass gain of the vessel corresponds to the non-volatile residue; dividing by the sample quantity yields the reported result.

Standard Operating Procedure

  • Vessel preparation: clean a quartz, platinum, or borosilicate-glass evaporating dish, dry it at 105 °C ± 2 °C, cool in a desiccator for approximately 30 minutes, and weigh on an analytical balance readable to 0.1 mg; repeat the cycle until constant mass is achieved, with successive weights differing by no more than about 0.2 to 0.3 mg.
  • Sample addition: accurately pipette the volume stipulated by the product standard, commonly 50 mL to 200 mL, or weigh the sample directly for viscous or low-volatility materials.
  • Evaporation: place the dish on a boiling water bath inside a fume hood or clean enclosure and allow slow, dust-free evaporation without splashing or boiling over.
  • Drying and weighing: transfer the dish to a forced-convection oven at 105 °C ± 2 °C or the monograph-specified temperature, dry for the prescribed time, cool, and weigh; repeat heating, cooling, and weighing cycles until constant mass is demonstrated.
  • Blank and replication: run vessel blanks and duplicate determinations to control environmental deposition and weighing drift.

Calculation

As a mass concentration, the non-volatile matter is calculated as X = (m1 − m0) / V, where m1 is the mass of the dish plus residue, m0 is the mass of the empty dish, and V is the sample volume. As a mass fraction, w = (m1 − m0) / m × 100 %. Blank-corrected results are compared with specification limits, for example no more than 0.0005 percent (5 ppm) for premium-grade ethanol and commonly no more than 0.001 percent for many analytical-grade solvents.

Equipment and Instrumentation

  • Analytical balance with 0.1 mg readability, periodically calibrated
  • Quartz, platinum, or borosilicate-glass evaporating dishes
  • Boiling water bath with stable temperature control
  • Forced-convection drying oven regulated to within ± 2 °C
  • Desiccator with fresh desiccant and Class A volumetric pipettes
  • Clean bench or laminar-flow enclosure for trace-level determinations

Method Notes for Difficult Matrices

Hydrofluoric acid and ammonium fluoride solutions attack glass and quartz, so platinum dishes are mandatory. Highly volatile samples should be pre-chilled and transferred quickly to avoid mass loss before weighing. Oxidizing acids must be evaporated slowly under efficient ventilation, and samples prone to creeping require pre-conditioned, slightly inclined dishes to prevent residue loss over the rim.

Applications

  • Lot-release quality control at chemical reagent manufacturing plants, verifying compliance with guaranteed-grade and analytical-grade specifications
  • Incoming inspection by pharmaceutical, biotechnology, and food testing laboratories that rely on reagent purity for data integrity
  • Qualification of process chemicals for semiconductor, display panel, and photovoltaic fabrication, where residue directly affects yield
  • Solvent quality assessment for paints, coatings, inks, and adhesives in accordance with ASTM D1353
  • Verification of solvent recovery and recycling streams to confirm removal of dissolved solids and high-boiling contaminants
  • Arbitration testing, regulatory compliance, and research studies requiring independent, accredited third-party laboratories

FAQ

Question: What does an elevated non-volatile matter result indicate?

Answer: An elevated result points to residual inorganic salts, dissolved or suspended solids, high-boiling organic impurities, excess stabilizer, or leachables from containers and transfer lines. It may reflect incomplete purification, cross-contamination, solvent degradation, or poor storage practices.

Question: What is the practical detection capability of the method?

Answer: With a 100 mL sample and a 0.1 mg balance, the gravimetric method reliably quantifies residues down to approximately 1 to 2 mg/L. Lower levels require larger sample volumes, clean-room handling, or instrumental techniques such as ICP-MS and ion chromatography.

Question: Why is 105 °C the default drying temperature?

Answer: 105 °C ± 2 °C removes residual moisture and solvent traces without decomposing most residues, giving stable and reproducible weights. Certain monographs specify alternative temperatures or drying times when the chemical nature of the expected residue requires them.

Question: How does non-volatile matter differ from ignition residue?

Answer: Non-volatile matter is dried without ignition and therefore captures both organic and inorganic residue, whereas ignition residue per GB/T 9741 involves carbonization and ashing, leaving only the inorganic ash. The two values together help distinguish organic contamination from inorganic contamination.

Question: Are blank corrections necessary?

Answer: Yes. Vessel and environmental blanks are essential, especially at trace levels, because airborne dust and balance-room contamination can contribute a measurable mass. Only blank-corrected results should be compared against specification limits.