Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

The plastic color difference determination test quantifies the deviation in color between a plastic sample and a reference standard using instrumental colorimetry, with results expressed primarily as ΔE*ab in the CIELAB color space. This test is essential for verifying batch-to-batch color consistency of plastic materials and finished parts, and it is routinely performed by accredited third-party laboratories in accordance with standards such as ISO 11664, ASTM D2244, ASTM D6290, and GB/T 3979. This article explains the testing principles, sample requirements, measured parameters, procedures, and typical industrial applications of color difference testing for plastics.

Technical Overview

Color difference determination is an instrumental method used to quantify how far the color of a plastic specimen deviates from an agreed reference color, such as a master standard plaque, an approved golden sample, or a specified color coordinate. Because human visual judgment of color is subjective and influenced by lighting, observer, and surface finish, objective colorimetric measurement is required for quality control, supplier approval, and dispute resolution.

The most widely used evaluation system is the CIELAB (L*a*b*) color space defined in ISO 11664-4. In this system, color is described by three coordinates:

  • L* – lightness, ranging from 0 (perfect black) to 100 (perfect white);
  • a* – the red–green axis, where positive values indicate redness and negative values indicate greenness;
  • b* – the yellow–blue axis, where positive values indicate yellowness and negative values indicate blueness.

The total color difference between the sample and the reference is calculated as:

ΔE*ab = √[(ΔL*)² + (Δa*)² + (Δb*)²]

where ΔL*, Δa*, and Δb* are the differences in each coordinate. The individual coordinate differences are equally important because they indicate the direction of the deviation (darker/lighter, redder/greener, yellower/bluer), which is critical for color correction in production. For improved correlation with human perception, particularly for saturated colors, the CIEDE2000 color difference formula (ΔE00), standardized in ISO/CIE 11664-6, is increasingly specified as an alternative to ΔE*ab.

Measurements are typically performed under CIE standard illuminant D65 with either the 2° or 10° standard observer, most commonly the 10° observer. Under controlled viewing conditions, a ΔE*ab of approximately 1.0 is generally the threshold of perceptibility; industrial acceptance tolerances typically range from 0.5 to 2.0 depending on the application, color, and surface texture.

Testing Samples

Injection-Molded Plaques and Panels

Flat, opaque plaques molded under controlled conditions are the preferred laboratory sample form. Standardized plaques eliminate surface geometry effects and allow direct comparison between material lots, masterbatch dosages, or processing parameters.

Plastic Pellets and Granules

Raw pellets can be measured directly using a special pellet holder or glass vessel, typically with a large-aperture sphere spectrophotometer. This approach is standardized in ASTM D6290 and is useful for incoming material inspection, although pellet color only approximates the color of the molded part.

Films and Sheets

Transparent and translucent films are measured in transmission or reflectance mode, often with backing material of defined color to standardize the measurement. Multiple layers may be stacked to achieve opacity for reflectance measurement.

Finished Parts and Components

End products such as housings, trims, caps, and panels can be measured directly on flat, representative areas, provided the measurement aperture can make full contact with the surface.

Powders and Compounds

Rotational molding powders, PVC compounds, and masterbatch are pressed into plaques or measured in a powder cell with a defined packing method to ensure repeatability.

Testing Items

Typical parameters determined in a plastic color difference test include:

  • Tristimulus color coordinates: L*, a*, b* (and X, Y, Z values where required);
  • Coordinate differences: ΔL*, Δa*, Δb*;
  • Total color difference: ΔE*ab (CIELAB) and/or ΔE00 (CIEDE2000);
  • Yellowness index (YI) per ASTM E313 and ISO 17223, used to detect thermal or UV degradation;
  • Whiteness index per ASTM E313 or the CIE whiteness formula;
  • Color change (ΔE) after heat aging, UV or xenon-arc weathering per ISO 4892, or chemical exposure;
  • Grey scale rating for color change per ISO 105-A02 / GB/T 250, where a visual rating method is required;
  • Metamerism index, where samples must match under multiple illuminants.

Testing Methods

Instrumental Spectrophotometric Method

The reference method uses a calibrated spectrophotometer to measure the spectral reflectance (or transmittance) of the specimen from approximately 360 nm to 780 nm. Reflectance data are weighted by the selected illuminant and observer functions, per ASTM E1164 and ASTM E308, to compute the tristimulus values and L*a*b* coordinates. Common measurement geometries include:

  • d/8 sphere geometry – diffuse illumination with 8° viewing; may be operated with the specular component included (SCI) or excluded (SCE);
  • 45/0 (or 0/45) geometry – directional illumination and viewing, which correlates closely with visual assessment under normal viewing conditions.

SCI measurements include the specular reflection and therefore represent the "body color" independent of surface gloss or texture, whereas SCE measurements exclude the specular component and better simulate what an observer sees on glossy surfaces. The instrument is calibrated before use with a white tile and a black trap (or light trap) traceable to national standards.

Color Difference Calculation

Color tolerances and differences are calculated from the measured coordinates in accordance with ASTM D2244, ISO 11664, GB/T 3979, and GB/T 11186.3. A minimum of three readings at different locations on the specimen are averaged, and the mean values are compared with those of the reference standard measured under identical conditions.

Visual Comparison Method

Where visual appraisal is required, samples are compared against the standard in a standardized light booth under D65, A, and shop-light sources in accordance with ASTM D1729 and ISO 3668. For color change after aging or exposure, the grey scale method of ISO 105-A02 / GB/T 250 assigns a rating from grade 1 (large change) to grade 5 (no change).

Test Conditions and Preparation

Specimens are conditioned prior to measurement, typically at 23 ± 2 °C and 50 ± 5 % relative humidity per ISO 291 / GB/T 2918. Samples must be clean, free of fingerprints and scratches, opaque (or consistently backed), and large enough to cover the measurement aperture. Curved or textured parts should be measured at identical positions on both sample and reference to minimize geometric bias.

Principal Reference Standards

  • ISO 11664 series – Colorimetry (illuminants, observers, CIELAB, CIEDE2000);
  • ASTM D2244 – Calculation of color tolerances and color differences from instrumentally measured coordinates;
  • ASTM E308 and ASTM E1164 – Computing object colors and obtaining spectrophotometric data;
  • ASTM D6290 – Color determination of plastic pellets;
  • ASTM E313 – Yellowness and whiteness indices;
  • ISO 17223 – Determination of yellowness index of plastics;
  • GB/T 3979 and GB/T 11186 – Chinese national standards for object color measurement and color difference calculation;
  • JIS Z 8722 – Measurement of object color reflectance (Japanese standard).

Applications

Color difference determination is applied across the entire plastics value chain, from resin and masterbatch production to finished part acceptance. Typical applications include:

  • Incoming inspection of resins, compounds, and color masterbatch against approved standards;
  • Batch-to-batch color consistency control in injection molding, extrusion, blow molding, and thermoforming;
  • Automotive interior and exterior trim, where tight tolerances (often ΔE*ab ≤ 0.5–1.0) are required for multi-part assemblies;
  • Consumer electronics housings, household appliances, and cosmetic packaging, where brand color identity is critical;
  • Evaluation of color stability after UV weathering, heat aging, sterilization, or chemical exposure;
  • Root-cause analysis of color defects (for example, masterbatch dosage, residence time degradation, or contamination);
  • Sign-off testing and dispute resolution by accredited third-party laboratories.

FAQ

What is a color difference test for plastics?

It is an instrumental test that measures the color coordinates of a plastic sample and compares them with a reference standard, expressing the deviation as ΔL*, Δa*, Δb*, and the total color difference ΔE*ab (or ΔE00) calculated per ASTM D2244, ISO 11664, and GB/T 3979.

What ΔE*ab value is considered acceptable?

It depends on the application and the color. As a guideline, ΔE*ab below 0.5 is considered an excellent match, 0.5 to 1.0 is typical for automotive and premium products, 1.0 to 2.0 is common for general consumer goods, and values above 2.0 to 3.0 are usually visible to the naked eye and rejected. Contractual tolerance limits should always be agreed between supplier and customer.

Can plastic pellets be tested directly?

Yes. Pellets are measured in a dedicated pellet holder or glass vessel with a large-aperture instrument in accordance with ASTM D6290. However, pellet color is influenced by scatter and only approximates the final molded-part color, so molded plaque confirmation is recommended for critical colors.

What is the difference between SCI and SCE measurement?

SCI (specular component included) excludes the influence of surface gloss and texture, representing the intrinsic material color, while SCE (specular component excluded) better reflects how the part appears visually. The required mode must be specified because the two modes can yield different ΔE values on the same textured or glossy part.

How should samples be prepared for the test?

Samples should be conditioned at 23 ± 2 °C and 50 ± 5 % relative humidity, cleaned, flat and opaque (or uniformly backed), and at least as large as the measurement aperture. Sample and reference must be measured under identical illuminant, observer, geometry, and aperture settings.

Why is color difference testing important for plastics?

Color deviation is often the first visible sign of process drift, material degradation, or masterbatch error. Quantitative ΔE data enables objective quality control, reduces subjective disputes over color matching, and supports validation of color durability after weathering or thermal aging in regulated industries.